| Product name | Continuous Particulate Monitor with X-ray Fluorescence |
|---|---|
| Model | PX-375 |
| Measured object | Particulate matter (PM10, PM2.5, TSP) |
| Measurement content | Particulate mass concentration and element concentration |
Common
| Flow rate | 16.7L/min |
|---|---|
| Sampling pump | Linear drive system, externally installed |
| Filter tape | None-woven PTFE fabric filter |
| Spot tape interval | 20/25/50/100mm selectable |
| Filter tape replacement interval | Approx. 1 month (In case of 100mm spot interval) |
| Ambient operation temperature | 10˚C~30˚C |
| Relative humidity | 0~80% RH noncondensing |
| Altitude | 1000m or less |
| Power supply | AC100V~240V ±10%, 50/60Hz±1% |
| Power consumption | Approx. 400VA |
| External dimension | 430mm(W)×550mm(D)×285mm(H) (without sampling pipe and measurement head) |
| Weight | Approx. 40kg (Without sampling pipe and measurement head) |
| Data output | CSV file (Average PM mass and elemental concentration) |
| External connection | EthernetTM, USB, RS-232C* (option) |
*Please consult about communication and instrument composition separately.
Mass analyzer unit
| Measurement method | Beta-ray attenuation |
|---|---|
| PM10 | US EPA Louvered PM10 Inlet |
| PM2.5 | BGI VSCCTM Cyclone |
| TSP | TSP Inlet |
| Measurement range | 0~200/500/1000μg/m3 |
| Repeatability | ±2% (against reference foil value) |
| Span drift | ±3% (24hours) |
| Lowest detection limit (2σ) | ±4μg/m3 (24hours) |
| Sampling and measurement cycle | 0.5/1/2/3/4/6/8/12/24 hours |
Element analyzer unit
| Measurement method | Energy dispersive X-ray spectroscopy |
|---|---|
| Detectable elements | See Table 2 "Detectable Elements". Standard parameter is S, Ti, Cr, Mn, Ni, Cu, Zn, Pb, Al, Si, K, Ca, V, Fe, As. |
| Primary X-ray filter | Automatic switching for light metals/heavy metals |
| Tube voltage | Automatic switching for 15kV/50kV |
| Detector | SDD (Silicon Drift Detector) |
| Sample image | CMOS camera |
| Lowest detection limit (2σ) | Recommended EPA Method IO 3.3 See Table 1 "Lowest Detection Limit (Example)" |
| Measurement range | Up to measurement time |
| Analysis time | 1000s (16.6 min) as standard 100 / 200 / 500 / 1000 / 2000 / 5000 / 10000s selectable |
| Calibration material for X-ray intensity for standard parameter | NIST SRM 2783, other materials (option) |
| Safety functions for X-ray | Internal lock system Key switch X-ray indication light |