System Function
Fully automated, discrete, random access, with STAT sample priority capability
Throughput: 400 photometric tests/hour, up to 240 tests/hour for ISE
Measuring principles: Absorbance Photometry, Turbidimetry
Methodology: End-point, Fixed-time, Kinetic, ISE (Optional)
Single/Dual/Triple/Quadruple reagent chemistries, Monochromatic/Bichromatic
Programming: User defined profiles and calculation
Reaction System
Reaction rotor: Rotating tray, 90 cuvettes with automatic washin
Cuvette: Optical length 5mm
Reaction volume: 120-360μl
Operating temperature: 37°C
Temperature fluctuation: ±0.1°C
Mixing system: 2 independent mixers
Sample Handling
Sample tray: 90 positions for primary or secondary tubes and sample cups
Sample volume: 1.5-45 μl, step by 0.1μl
Sample probe: liquid level detection, clot detection and collision p- rotection
Probe cleaning: Interior and exterior automatic probe washing carry-over < 0.05%
Automatic sample dilution: Pre-dilution and post dilution
Dilution with ratio up to 1: 150
Dilution vessel: Permanent Cuvette
Optical System
Light Source: Halogen-tungsten lam
Photometer: Reversed optics, grating photometry
Wavelength: 340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm, 605nm, 660nm, 700nm, 740nm, 800nm
Absorbance range: 0-3.3Abs (10mm conversion)
Resolution: 0.0001Abs
Internal Bar Code Reader
Sample/ Reagent barcode reading including Codabar, ITF (Interleaved Two of Five), code128, code39, UPC/EAN, Code93; Bi-directional LIS Interface transmission
Control and Calibration
Calibration mode: Linear (one-point, two-point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline, exponential, Polynomial, Parabola
Control rules: Westgard multi-rule, Levy-Jennings, Cumulative sum check, twin plot
ISE Module (optional)
Optional selection of K+, Na+, Cl-
Throughput: Up to 240 tests per hour
Operation Unit
Interface: RS-232, Network Port, USB/ parallel port
Reagent Handling
Reagent tray: 80 positions in refrigerated compartment
Reagent volume: (2-10°C) R1: 45- 350 ul, R2-R4 :10 - 350ul
Reagent probe: Liquid level detection, collision protection and inventory check
Probe cleaning: interior and exterior automatic probe washin
Working Conditions
Power Supply: 200-240V, 50/60Hz, 1500V or 110-130V, 60Hz, 1500V
Temperature: 15-30°C
Humidity: 35-85%
Water consumption: cS20 L/hour, Deionized water
Dimension: 46in x 28in x 45in (W x D x H)
Weight: 661lb