InverTau Specifications
Modes | Single point, raster, bi-directional (xy) |
Scan speed | Up to 6 fps (bi-directional) at 512x512 |
Image resolution | 32x32 to 4k x 4k |
Digital zoom | 1FoV to 0.1FoV |
Attachment | To side port of inverted microscope (C-mount) |
Control software | EzTime™ Image software |
Optical ports | 2 on excitation, 2 on emission |
Attenuation filters | Steps from 0-6 OD ND |
Dichroic filters | 400 nm, 450 nm, 500 nm, 550 nm, 600 nm, 730 SP, 50/50 BS |
Emission filters | 400BP, 450BP, 500BP, 550BP, 600BP, 440LP, 475LP, 530LP, 570LP, 630LP |
Pinhole | 10, 25, 50, 100, 200 μm (+ closed and fully open) |
Photon streaming | Stream to HDF5 |
Phasor analysis | Switchable with histogram |
Data analysis | Global up to 5 exponentials |
Key FiPho Electronics Specifications for FLIM
Lifetimes from | <50 ps |
Deadtime | 5 ns |
TCSPC bin width | <15 ps |
Histogram Bin Depth | 8, 16 and 32-bit |
Independent Stops | 1 to 4 |
Maximum Start Rate | 100 MHz |
Maximum Rate/Stop Channel | 10 Mcps |
Operating Mode | Automatic Forward Timing |
Streaming Mode | Photon Streaming (Time-Tag) |
Acquisition and Analysis Macro Scripting | Yes |
PC Interface | USB 3.0> |
Software | EzTime™ Image |
View Hybrid Picosecond Photon Detector (HPPD) Series Specifications