
| Measuring Principles | Batch sampling: Filter gravimetric method Real-time PM measurement : Diffusion charging method | 
| Measuring Ranges (Real-time PM sensor) | 0 to 2500 mm/cm3 (Concentration after dilution) | 
| Dilution Method | Partial flow dilution method (Proportional sampling mode/Fixed dilution ratio mode) | 
| Signal for Gas Sampling | Exhaust gas flow rate: Isolated analog signal (0-10 V) | 
| Flow Rate of Diluted Gas | 10 to 15 L/min (Under the conditions of 20 deg.C and 101.3 kPa) | 
| Filter Holder | Diameter 47 mm, Effective diameter 26.2 mm (Without backup filter) | 
| Particle Pre-classifier (Cyclone) | 99% penetration of particle size of less than 1 μm 50% cut off point: particle size between 2.5 and 10 μm | 
| Power Requirements | DC 24 V | 
| Power Consumption | Max. 0.6 kVA | 
| Dimensions | 350(W) x 470(D) x 460(H) mm (Excluding controller and protruding objects) | 
| Mass | Approx. 45 kg (Excluding controller) | 
| Operating Conditions | Temperature: 0 to 40 deg.C (During operation) Relative humidity: less than 80% Altitude: 0 to 2000 m above sea level | 
System Configuration
Combination of PM measurement unit and GAS measurement unit


 EPA CFR part 1065 Compliance
 EPA CFR part 1065 Compliance


































